7. Documentation
7.1. Photography over Photo-/TV tube
If visual observation and photographic
photographs without changes are to be made,
then the use of the Photo-/TV tube offers itself. A
photo adjustment and the type of camera
appropriate T2 – adapter is needed. There are
different photo adjustments available: 1x ; 1,6x
and 3,2x.
It is better to insert into the adjustable eyepiece a
cross-line plate to see which part of the object
(picture) will be shown on the film and to
suppress the individual accommodation.
The Photo-/TV tube is mounted on the
intermediate tube and the photo adjustment on
the upper exit of the Photo-/TV tube itself. The
objective is removed from the camera and the
T2-adapter will be mounted there.
The photo adjustment cannot be adjusted,
because it is so balanced that after the normal
alignment of the microscope also the picture
appears sharp on the film level.
7.2 Digital photography
The connection of digital single lens reflex
cameras is done by a sensor fitted photo
adaptation and camera suited T2 adaptation
ring. Three different photo adaptation 1,0x;
1,6x; 3,2x are available.
For a maximum object field following optimal
combinations between sensor size an photo
adaptation are recommend:
Full frame 24,0mm x 36,0 mm 3,2x
APS-C 14,8mm x 22,2mm 1,6x
Four Thirds 13,5mm x 18,0mm 1,0x
pic.13: photo adaptations 3,2x; 1,6x; 1,0x